Differential Pressure Sensor
Digital differential pressure sensor for cleanrooms, labs, and HVAC monitoring with automated calibration.
Technical Specifications
Detailed specifications for Differential Pressure Sensor
The Eupry differential pressure sensor (P4P2) monitors pressure differentials in cleanrooms, laboratories, operating rooms, and ventilation systems.
The sensor mounts on the cleanroom wall and connects to a Eupry data logger through a 4 mm tube routed between the two rooms whose pressure differential is being measured. No drilling, no cabling, no panel penetrations. The cleanroom envelope stays sealed throughout install.
Calibration is automated. A new calibrated sensor ships before your current certificate expires. Swap it in, and the new ISO 17025 certificate appears in the monitoring software automatically.
The sensor supports cleanroom qualification, HVAC validation, laboratory monitoring, and operating room pressure cascade verification.
3 steps to operational
The Eupry DP sensor is a small, battery-powered wireless unit – about the size of a deck of cards. It transmits data to the Eupry platform via the hub already in your facility.
- Mount the sensor on the cleanroom wall. No drilling into the wall structure.
- Run a 4.8 mm tube between the two rooms (through an existing port or under a door). That is the only physical connection.
- The sensor is online. Readings transmit wirelessly to the Eupry platform.
No construction, no facility downtime, no re-qualification of the room.
Compatible products & add-ons
Get a free product data sheet
Ready to get started?
See how Eupry handles monitoring, calibration, and validation in one platform.